Hochdruckreaktor, Autoklav, Magnetrührdurchführung, Hydrierung

andorra

High pressure reactor andorra

"andorra" comes from the small autoclave family - a development in autoclave construction based on many years of experience and perfected. Reactor "andorra" has the technology that is otherwise only found in large reactors: Seals have been severely limited, thanks to the prudent design to merge the reactor lid with the core piece of the magnetic stirrer drive and the magnetic stirrer drive in turn has, in addition to the essential valve connections, a pressure measurement and the pressure protection.

Specifications

  • Nominal volume ml: 60, 100
  • Working volume ml: 20-40, 40-60
  • Working pressure bar: 200, 280, 480
  • Working temperature °C: 200, 250, 300
  • Material WNr. 1.4435 at 200 bar - WNr. 2.4602 at 250 bar - WNr. 1.4980 at 280 bar
  • Union nut as lock M48 x1.5 mm, width across flats SW55
  • Double sealed: O-ring and medium seal metal to metal (cutting ring).
  • Reactor cover resp. magnetic stirrer drive with a torque of 20 Ncm and integrated micromotor 24V/DC
  • Output speed with ball bearing 0-1'300 rpm, with plain bearing 0-1'200 rpm.
  • Regulating shut-off valve for gas supply
  • Regulating shut-off valve for pressure relief
  • Regulating shut-off valve with immersion tube for sampling
  • Pressure gauge in chemical version
  • Bursting disc holder equipped with bursting disc
  • Immersion tube for thermocouple with ø 3 mm (without thermocouple)
  • Temperature sensor Pt100 4-wire or type K suitable
  • Stirring element: gas supply stirrer ø 18 x 64 mm

Variants/special features

  • Pressure stages: 200, 250 or 480 bar depending on the choice of material.
  • Temperature stages °C: 200, 250, 300 depending on the choice of material.
  • Material for wetted parts WNr.: 1.4435 (AISI 316L), 2.4602 (HC22) and 1.4980 (AISI 660) as standard. Versions in Hastelloy C276, B3 or titanium size 2 on request.
  • The autoclave vessel is double-sealed: once in the upper area by an O-ring made of Viton, EPDM or Kalrez and at the bottom via a metal-to-metal medium seal (cutting ring).
  • Temperature sensor Pt1000, type K or type N in the immersion tube for measuring the medium temperature (not included in delivery).
  • By opening the locking nut SW12 mm, the output shaft can be removed for cleaning the inside or bearing replacement. Either ball bearings or plain bearings are used, depending on the medium or material used.

A real team player!

"andorra" is a real team player when it comes to multi-versions of four, eight or more autoclaves in series. The small autoclaves can easily be used in multiple reactions, where they are heated, cooled or controlled independently.

Article No Vol. mlPres PS barTemp. TS C°Pres PA barTemp. PA C°Mat. WNr.Sensor / HeatingDrive / VersionL x W x H mm
129908 602502252002001.4435Type K3 regulating valves185 x 185 x 536
129919 602502252002001.44351 x Pt100, 4 wires3 regulating valves185 x 185 x 536
132348 603502752802502.4602Type K3 regulating valves185 x 227 x 536
132353 603502752802502.46021 x Pt100, 4 wires3 regulating valves185 x 227 x 536
132346 606003254803001.4980Type K3 regulating valves185 x 227 x 536
132354 606003254803001.49801 x Pt100, 4 wires3 regulating valves185 x 227 x 536
129910 1002502252002001.4435Type K3 regulating valves185 x 185 x 575
129920 1002502252002001.44351 x Pt100, 4 wires3 regulating valves185 x 185 x 575
132349 1003502752802502.4602Type K3 regulating valves185 x 227 x 575
132352 1003502752802502.46021 x Pt100, 4 wires3 regulating valves185 x 227 x 575
132342 1006003254803001.4980Type K3 regulating valves185 x 227 x 575
132355 1006003254803001.49801 x Pt100, 4 wires3 regulating valves185 x 227 x 575
* For a direct quote request, click the corresponding article number.